Abstract/Description
Multi-needle Langmuir probe is a fairly new instrument technique that has been flown on several recent sounding rockets and is slated to fly on a subset of QB50 CubeSat constellation. This paper takes a fundamental look into the data analysis procedures used for this instrument to derive absolute electron density. Our calculations suggest that while the technique remains promising, the current data analysis procedures could easily result in errors of 50% or more. We present a simple data analysis adjustment that can reduce errors by at least a factor of five in typical operation.
Publication Title
Review of Scientific Instruments
DOI
https://doi.org/10.1063/1.5022820
Publisher
AIP Publishing
Scholarly Commons Citation
Barjatya, A., & Merritt, W. (2018). Error Analysis of Multi-Needle Langmuir Probe Measurement Technique. Review of Scientific Instruments, 89(4). https://doi.org/10.1063/1.5022820